Diagnostic accuracy of smartphone ECG in primary care


  • Mary Corcoran
  • Univadis Medical News
Access to the full content of this site is available only to registered healthcare professionals. Access to the full content of this site is available only to registered healthcare professionals.

New research suggests that a handheld, smartphone-enabled, single-lead electrocardiogram (ECG) device could provide physicians with a viable alternative to traditional ECGs to detect cardiac arrhythmias in patients presenting with non-acute cardiac concerns in primary care. 

In a multi-centre validation study of the diagnostic accuracy of the one-lead device (KardiaMobile), blinded cardiologists assessed data from 214 Dutch primary care patients collected simultaneously from the one-lead device with an integrated algorithm for atrial fibrillation (AF) against a 12-lead device. 

The 12-lead ECG diagnosed AF/atrial flutter (AFL), any rhythm abnormality and any conduction abnormality in 23, 44 and 28 patients, respectively. 

The one-lead ECG, as assessed by cardiologists, had a sensitivity and specificity for AF/AFL of 100 per cent for both. The algorithm correctly identified 87 per cent of AF cases and 97.9 per cent of non-AF cases.

However, the one-lead ECG was less robust at categorising rhythm abnormality (sensitivity: 90.9%; specificity: 93.5%) and conduction abnormality (sensitivity: 46.4%; specificity: 100%). 

“Our primary care-based study provides important insights for physicians who are in need of a point-of-care ECG device that can lower the logistical threshold for performing ECG to improve diagnostic gain,” the authors said. The findings are published in the Annals of Family Medicine